85DKD-85DX-6050 (optional)
Burster
DAkkS calibration certificate according to DIN EN ISO 376 for
Force sensors or measuring chains in the direction of pressure,
10 force levels evenly over the measuring range
distributed (10% steps).
Increasing and decreasing measuring force in different
Mounting positions.
For measuring range end values 100 N to 50 kN.
You can obtain your calibration certificate from:
www.burster.de/Kalibrierung
*********************************
If there is a statement of conformity in the calibration certificate
for compliance with error limits is specified, so
we only make the determination below
Consideration of the expanded measurement uncertainty
(Confidence level 95%), according to ILAC-G8: 09/2019,
2019 Apendix B Example 2.
This means that compliance with
Data sheet information of the calibration item only
can be confirmed provided the measurement deviation
plus the amount of the extended
Measurement uncertainty within the permissible
Specification lies.
Statements of conformity for device classes that are normative
are regulated (DIN 51309 and DIN EN ISO 376)
we meet according to the procedures mentioned there.
*********************************